Description
Maquet Datascope CS300 IABP with IntelliSense
The Maquet Datascope CS300 Intra-Aortic Balloon Pump (IABP) with IntelliSense combines fiber-optic speed with automatic in vivo calibration, delivering faster therapy, quicker signal acquisition, and rapid adaptation to changes in heart rate and rhythm. It is the only fiber-optic IABP and catheter system that automatically calibrates within the patient after insertion and re-calibrates every two hours—or sooner if patient or environmental conditions change.
Key Features:
-
Augmentation and support during diastole for improved cardiac output
-
Faster inflation and deflation speeds for enhanced hemodynamic support
-
True one-button start-up for rapid operation
-
Automatic in vivo calibration of the fiber-optic pressure sensor with re-calibration every 2 hours or sooner as needed
-
Automatic lead and trigger evaluation for optimal timing and performance
-
Automatic timing optimization to maximize therapy efficiency
Specifications:
Dimensions & Weight:
-
On Cart: 43.1” H × 22.3” D × 16.8” W (109 × 56.6 × 42.7 cm)
-
Off Cart: 26.9” H × 20.5” D × 10.8” W (68.3 × 52.1 × 27.4 cm)
-
Console Weight: 84.8 lbs (38.4 kg) nominal (includes S.D., CRM, and He tank)
-
Monitor: 9.5 lbs (4.3 kg) nominal
-
Hospital Cart: 52.4 lbs (23.8 kg) nominal
-
Internal Battery: 34 lbs (15.4 kg) nominal
-
Storage Bag: 4.6 lbs (2.1 kg) nominal (includes doppler and holder)
Display:
-
Color TFT LCD for high-visibility monitoring
Trigger Options:
-
ECG Trigger: Automatically adjusts threshold for high sensitivity and selectivity of R-wave detection (Min: 120 µV ± 20 µV at normal gain; 40 µV at max gain)
-
Pressure Trigger: Default threshold automatically set at 38% of systolic pulse height (min: 7 mmHg); manual adjustable range 7–30 mmHg ± 3 mmHg
-
Pacer Triggers:
-
A Trigger: R-wave detection with 100 ms pacer blanking
-
V/A-V Trigger: V pacer up to 185 bpm, A-V pacer up to 125 bpm with intervals 80–224 ms
-
Pressure Output:
-
Electrically isolated, range 0–+300 mmHg
ECG:
-
Leads: Auto Mode – I, II, III, External; Semi-Auto – I, II, III, AVR, AVL, AVF, V, External (12-lead compatible)
-
Gain: 1 V per 1 mV input ±5%, waveform automatically scaled to display window
The CS300 provides rapid, reliable, and precise cardiac support, combining cutting-edge fiber-optic technology with fully automated calibration and timing features for optimal hemodynamic assistance.